• Product Search
    Enter a part number or product name to perform a quick search of results with your search term...
    Search Term:
    Results
    • Cart
    9
    9
    Pine AfterMath FAQ: AT-MIO-16E-10 DMA Error

    Pine AfterMath FAQ: AT-MIO-16E-10 DMA Error

    by | Jan 4, 2016 | Uncategorized

    [call_to_action color=”green”]

    AfterMath_NEW
          This FAQ applies to versions 1.0.2447 and greater.[/call_to_action]

    Attempting to perform an experiment from AfterMath on the Pine AFCBP1 bipotentiostat using the ISA-type National Instruments card (AT-MIO-16E-10) may result in the following error:

    Experiment failed: "<experiment_name_here>". Instrument failure.

    This problem can be caused by a lack of DMA resources. This can be confirmed using either AfterMath or the National Instruments Measurement & Automation Explorer as shown below:

    atmio16e10-failure-08.jpg

    atmio16e10-failure-07.jpg

    atmio16e10-failure-09.jpg

    The original cause of this failure occurs during the installation of the National Instruments interface board device drivers, and can also be seen in the Windows Device Manager:

    atmio16e10-failure-01.jpg

    atmio16e10-failure-02.jpg

    The problem can be fixed by using the Windows “Device Manager” as shown below.

    See also the National Instruments website article about this problem here

    atmio16e10-failure-03.jpg

    atmio16e10-failure-04.jpg

    atmio16e10-failure-05.jpg

    atmio16e10-failure-06.jpg