• Product Search
    Enter a part number or product name to perform a quick search of results with your search term...
    Search Term:
    Results
    • Cart
    9
    9
    WaveNow User: Cell Port

    WaveNow User: Cell Port

    by | Jan 4, 2016 | Uncategorized

    [call_to_action color = “blue”] wavenow_tiny.jpg
    This article is part of the Pine WaveNow USB Potentiostat User's Guide[/call_to_action]


    [toggle title = “Cell Port”]

    The WaveNow USB Potentiostat is connected to an external electrochemical cell through a 15-pin cell port located on the side of the instrument. This port is a female HD-15 connector and has the same appearance as a VGA video connector on a personal computer.

    wavenow_cell_connection_port.jpg


    Pine offers several products designed to mate with the cell port on the WaveNow USB Potentiostat:

    • WaveNow/WaveNano Shielded Cell Cable Kit (part number AKCABLE5)
    • WaveNow/WaveNano Student Cell Cable for use with the Student Voltammetry Cell (part number RRTPE04)
    • Universal Dummy Cell for troubleshooting the potentiostat (part number AB01DUM1)
    • WaveNow Universal Cell Cable Kit (part number AKCABLE3) (discontinued)

    Note that some of the items listed above are sold separately.

    [/toggle]
    [toggle title = “Pinout Information”]

    Those users wishing to build their own cell cable for use with the WaveNow potentiostat should consult the pinout information listed in the table below. The color codes in the table refer to the standard Pine color code for electrochemical cell connections.

    wavenow_db15_pinout.jpg
    Pin
    1
    2
    3
    4
    5
    6
    7
    8
    9
    10
    11
    12
    1
    13
    14
    15
    Description
    reference electrode
    (shield for the reference signal)
    working electrode sense
    (shield for the sense signal)
    working electrode
    counter electrode
    (used for calibration)
    (used for calibration)
    reserved
    (working electrode signal shield)
    analog ground
    analog ground
    (used for calibration
    analog ground
    analog ground
    Color Code
    white
     
    orange
     
    red
    green
     
     
     
     
    black
    black
     
    black
    black
    [/toggle] [toggle title = “Additional Resources”] [/toggle]